Constant temperature incubator use precautions

How to use the constant temperature incubator :

Discharge water from the water inlet (above the rear plate) into the stainless steel liner, the water level reaches 2CM below the overflow, then turn on the power, put the cultured items into the workroom, close the studio door, turn on the power switch, set the temperature (CNC device, error ≦ 0.5 ° C)

Constant temperature incubator operation and maintenance :
1. Clean the condenser regularly (every 3 months): For air-cooled cooling of the compressor, the condenser fan should be regularly checked and the condenser should be decontaminated and dusted to ensure its good ventilation and heat transfer performance; In the case of water-cooled cooling, in addition to ensuring the inlet pressure and inlet temperature, the corresponding flow rate must be ensured, and the inside of the condenser should be cleaned and descaled regularly to obtain its continuous heat transfer performance.
2, regular (every 3 months) cleaning evaporation (dehumidification): due to the different levels of cleanliness of the sample, under the effect of forced air circulation, evaporation (dehumidification) will condense a lot of small particles such as dust, should be carried out regularly Cleaning.
3. If the equipment needs to be relocated, it is best to carry out under the guidance of our company's technical personnel to avoid unnecessary damage or damage to the equipment.
4, circulating fan, condenser fan cleaning and calibration: similar to the cleaning evaporator, because the working environment of the test box is different, the circulating fan, condenser fan will condense a lot of small particles such as dust, should be cleaned regularly .
5. Since the value of the temperature (wet) box is generally high, we recommend that it be placed in a relatively benign temperature environment. Our empirical temperature value is 8 ° C ~ 23 ° C. For laboratories that do not have this condition, it must be equipped. Proper air conditioner (air-cooled) or cooling tower (water-cooled).
6. Waterway and humidifier cleaning: If the waterway is not smooth and the humidifier is fouled, it may cause the humidifier to dry and may damage the humidifier. Therefore, the waterway and humidifier must be cleaned regularly.
7, adhere to professional management. Units with conditions should send special personnel to the supplier factory for training and training to obtain more professional maintenance and repair experience and capabilities.
8. After each test is completed, set the temperature near the ambient temperature, work for about 30 minutes, then cut off the power supply and wipe the inner wall of the working chamber.
9, long-term shutdown is not used, the product should be energized every half month regularly, the power-on time is not less than 1 hour.